13
Step 2: Set the reminder function
To set the reminder:
On, move the slider to
Off, move the slider to
Step 3: When the reminder goes off
Your BETACONNECT will begin to beep and the top of the
blue light bar will flash to remind you when it is time to inject.
The flashing light reminder lasts for 1 hour.
The beep occurs every 15 minutes for 1 hour.
You can mute the beep and still have the flashing light
reminder by pressing the injection button or opening the
lid.
To cancel both the beep and the flashing light reminder,
press the on/off button.
Chapter 5: Troubleshooting
If you need assistance setting up, using, or maintaining your BETACONNECT autoinjector, talk with your BETA Nurse
or healthcare provider, or call BETAPLUS
®
at 1-800-788-1467 anytime.
Problem
Cause
Solution
Unable to turn the
BETACONNECT
autoinjector on
by pressing the on/off button
1. On/off button was not pressed
long enough.
1. Firmly press the on/off button for
more than 2 seconds.
2.
Battery is not charged.
2. Recharge the BETACONNECT
autoinjector.
3. Battery or electronic failure.
3.
Contact your BETA Nurse for a
replacement.
BETACONNECT autoinjector
turns off automatically
1. BETACONNECT autoinjector
will automatically turn off after
20 minutes of inactivity.
1. Press the on/off button to turn on.